Description of device parameters
Deltabar FMD71, FMD72
28
Navigation
Expert → System → Operator Code (021)
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a code to lock or unlock operation.
Entry
• To lock: Enter a number
¹
the release code (value range: 1 to 9999).
• To unlock: Enter the release code.
Note
The release code is "0" in the order configuration. Another release code can be defined in
the "Code Definition" parameter. If the user has forgotten the release code, the release code
can be visible by entering the number "5864".
Factory setting
0
Device tag (022)
Navigation
Expert → System → Instrument info → Device tag (022)
Write permission
Operators/Service engineers/Expert
Description
Enter the device tag, e.g. TAG number (max. 32 alphanumeric characters).
Factory setting
No entry or according to order specifications
Code Definition (023)
Navigation
Expert → System → Code Definition (023)
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a release code with which the device can be unlocked.
Options
A number from 0 to 9999
Factory setting
0
Level Selection (024)
Navigation
Expert → Measurement → Level → Level Selection (024)
Write permission
Operators/Service engineers/Expert
Description
Select the method for calculating the level
Summary of Contents for Deltabar FMD71