Description of device parameters
Deltabar FMD71, FMD72
62
Hi trim measured (276)
Navigation
Expert → Measurement → Sensor Trim LP → Hi Trim Measured (276)
Write permission
No write permissions. Parameter is read only.
Description
Displays the reference pressure present to be accepted for the upper calibration point.
Lo trim sensor (277)
Navigation
Expert → Measurement → Sensor Trim LP → Lo Trim Sensor (277)
Write permission
No write permissions. Parameter is read only.
Description
Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the lower calibration point.
Hi trim sensor (278)
Navigation
Expert → Measurement → Sensor Trim LP → Hi Trim Sensor (278)
Write permission
No write permissions. Parameter is read only.
Description
Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the upper calibration point.
Device type code (279)
Navigation
Expert → Communication → HART info → Device ID (279)
Write permission
No write permissions. Parameter is read only.
Description
Display of the numerical ID of the device
39
Sensor pressure LP (280)
Navigation
Expert → Measurement → Pressure → Sensor Press. LP (280)
Summary of Contents for Deltabar FMD71
Page 2: ......