Description of device parameters
Deltabar FMD71, FMD72
46
Options
• °C
• °F
• K
Note
The setting affects the unit for the "Sensor Temp." parameter.
Factory setting
°F
Density unit (127)
Navigation
Expert → Measurement → Level → Density Unit (127)
Write permission
Service engineers/Expert
Description
Displays the density unit. The measured pressure is converted to a height using the "Height
Unit", "Adjust Density" and "Process Density" parameters.
Options
• g/cm
3
• kg/m
3
• kg/dm
3
• lb/in
3
• lb/ft
3
Factory setting
g/cm
3
Electronics temp (128)
Navigation
Expert → Measurement → Basic setup → Electronics Temp. (128)
Write permission
No write permissions. Parameter is read only.
Description
Displays the serial number of the sensor module HP (11 alphanumeric characters).
Lo trim measured (129)
Navigation
Expert → Measurement → Sensor Trim HP → Lo Trim Measured (129)
Write permission
No write permissions. Parameter is read only.
Description
Displays the reference pressure present to be accepted for the lower calibration point.
Hi trim measured (130)