Operation with an operating menu
Deltabar FMD71, FMD72
8
User role
Submenu
Meaning/use
Diagnostics
Contains all the parameters that are needed to detect and analyze operating
errors. This submenu has the following structure:
•
Diagnostic list
Contains up to 10 currently pending error messages.
•
Event logbook
Contains the last 10 error messages (no longer pending).
•
Instrument info
Contains information for identifying the device.
•
Measured values
Contains all current measured values.
•
Simulation
Is used to simulate pressure, level, flow, current and alarm/warning.
•
Reset
•
Sensor LP
•
Sensor HP
Expert
Expert
Contains all the parameters of the device (including those already in one of the
submenus). The "Expert" submenu is structured by the function blocks of the
device. It thus contains the following submenus:
•
System
Contains all device parameters that neither affect measurement nor integration
into a distributed control system.
•
Measurement
Contains all parameters for configuring the measurement.
•
Output
Contains all parameters for configuring the current output.
•
Communication
Contains all parameters for configuring the HART interface.
•
Diagnostics
Contains all the parameters that are needed to detect and analyze operating
errors.
2.3
Operating options
2.3.1
Local operation
1
A0017650
1
Display and operating module with push buttons. Cover must be opened for operation.
2.4
Operating the device using onsite display (optional)
A 4-line liquid crystal display (LCD) is used for display and operation. The onsite display
shows measured values, dialog text as well as fault and notice messages in plain text,
thereby supporting the user in every stage of operation. The device display can be turned
in 90° steps. Depending on the installation position of the device, this makes it easy to
operate the device and read the measured value.
Functions:
Summary of Contents for Deltabar FMD71