Maintenance and care
multi EA 5100
142
CAUTION
Risk of injury from falling components
The user can be injured if the combustion furnace falls down during installation/re-
moval.
¡
Exercise extreme caution when installing/removing the combustion furnace.
The combustion furnace must be removed for transport.
}
Remove the combustion tube (→ "Maintenance of the multi-purpose
combustion tube"
126). Leave the furnace in the vertical position after
this.
}
Exit the multiWin program.
}
Switch off the basic module via the device switch and disconnect the
power plug from the socket. Switch off the gas supply.
}
Remove the top cover and the doors from the module.
}
Remove the left side panel: Disconnect the protective ground conductor.
Loosen the screws at the left side panel. Lift the side panel off and store
it safely.
}
Remove the hoses from their holders on the combustion furnace (see ar-
row).
}
Remove the protective grounding from the base plate of the combustion
furnace.
1
2
3
}
Remove the three plug-in connectors from their sockets:
– Flame sensor (1)
– Electrical connection of the combustion furnace (2). Press the gray
lever up slightly when doing this.
– Thermoelement (3) with colored cable