multi EA 5100
Transport and storage
169
WARNING
Risk of chemical burns due to acid and electrolyte solution return
Concentrated sulfuric acid is used in the Cl module 5100, phosphoric acid in the
TOC module 5100 and a slightly acidic electrolyte solution in the
S module 5100 coulometric.
A vacuum can be created in the analysis system when the combustion furnace is cooling.
This vacuum can lead to acid being drawn all the way into the auto-protection valve as-
sembly via connection hoses and lines.
¡
For the Cl module 5100: Only switch off the basic module and the gas supply when
the analysis system has cooled. The argon safety bypass on the auto-protection
valve assembly prevents a vacuum from being created during cooling for the chlo-
rine analysis branch. Alternatively: Remove the hose connection between the basic
module and the detection module before cooling.
¡
For the TOC module 5100 and the S module 5100 coulometric: Before shutting
down the basic module via the software, disconnection the hose connection to the
detection module.
}
Switch off the detection module via the power switch. Disconnect the power plug
from the power socket.
}
Shut off the gas supply.
}
Disconnect all connections on the rear of the detection module.
}
Seal the open gas connections with the ends of a short, fitting piece of hose to pre-
vent contamination during transport.
}
Remove loose components such as the absorber from the rear of the detection mod-
ule and package them individually.
}
If the detection module can be opened via the front door, remove all moving compo-
nents and package them individually. Observe the notes for individual detection
modules for this.
}
Carefully package the detection module and the accessories (cables, glass compo-
nents, hoses, clamps) in the original packaging.
}
Add a desiccant to the package to prevent moisture damage.
15.5.1 Notes for the transport of the Cl module 5100
WARNING
Risk of chemical burns
The concentrated sulfuric acid used as a drying agent and the acetic acid electrolyte so-
lution can cause severe chemical burns.
¡
Wear protective clothing when working on the sulfuric acid container and the mea-
suring cell.
¡
Observe all instructions and specifications in the safety data sheets.
}
Disconnect the electrodes (and the measuring cell) from the electrical connections on
the inside of the rear panel of the detection module.
}
Disconnect the measuring gas hose from the gas inlet tube/from the combined elec-
trode. Ensure that the seals of the PTFE screw connection are not lost during trans-
port.