Nitrogen analysis with the N module 5100
multi EA 5100
46
}
Insert the oxygen hose in the "O
2
" connector.
NOTICE! To release the hose, press the red ring into the connector and pull the
hose out of the connection.
}
Connect the detection module with the basic module via the interface:
"N-CLD" interface on the rear of the detection module
"N-CLD" on the rear of the basic module.
}
Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.
}
Connect an approx. 50 cm long hose to the "out" connection. Route the hose away
from the device.
ü
Connection of the detection module is complete.
The hose at the "out" connection prevents reaction gases from being sucked back into the
device by the absorber, disrupting the analysis. The hose does not have to be connected
to the laboratory exhaust unit.
1
2
Figure 24Connecting detection modules to the basic module
1 Detection module interfaces
2 Measuring gas outlet (for measur-
ing nitrogen, sulfur, carbon)