multi EA 5100
Carbon analysis with the C module 5100
85
11.2 Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡
Always connect the modules to power when they are switched off.
}
Place the detection module to the left of the basic module. In the event of a series of
detection modules: place the detection module to the left or right of the others.
}
Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
}
Connect the detection module with the basic module via the interface:
"C-NDIR" interface on the rear of the detection module
"C-NDIR" on the rear of the basic module.
}
Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.
}
Leave the "sample out" outlet unconnected or connect it to the measuring gas inlet of
the next detection module.
ü
Connection of the detection module is complete.
11.3 Operating the analysis system
}
Switch on the basic module and the detection module.
ü
The devices boot up. The status LED on the front of the basic module light up in
green after approx. 30 s.
ü
The LED on the front of the detection module flashes during the run-in period.
Depending on the detector, the run-in period can take up to 30 min. After this,
the LED will light up continuously. Starting a measurement is only then possible.
}
Open the gas supply and set the required gas pressure.
}
Switch on the PC.
}
Start the control and analysis software and login with your username and password.
}
Initialize the analysis system by clicking on [Initialize analyzer].
ü
The initialization and automatic detection of all connected components will be
carried out.
}
Ready the samples.
}
Activate a pre-existing method via the Method |Method - activate menu item.
}
Alternatively: Create a new method in the Method |Method - new menu. Select the
measurement parameter in the method. Release and activate the method.
}
Select Start |Start - Analysis in the menu.
}
Select an analysis group or create a new one and confirm via [OK].
}
Create an analysis sequence.
}
Enter the sample ID for all sample in the Name field.
}
Release all sequence lines
}
Confirm the entries with [OK].