multi EA 5100
Nitrogen analysis with the N module 5100
45
7.2
Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡
Always connect the modules to power when they are switched off.
NOTICE
The detection module is equipped with a pump that can disrupt the function of other
optical detection modules or lead to serious errors.
¡
Always connect the detection module as the last in any series of the detectors.
}
Install the detection module as the last in any series of detectors.
}
Install the absorber on the rear of the device:
– Fasten the two holding clamps with the supplied screws.
– First press the absorber into the upper holding clamp and then into the lower
one.
– Connect hose 6 to the absorber. Do not pull the hose out of the device!
1
2
3
Figure 23Absorber
1 Hose 6 connection
2 Holding clamp
3 Absorber
}
Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!