Specifications
multi EA 5100
174
Dosing speed (horizontal)
Liquids, EOX
1 to 10 µl/s
Recommended: 3 µl/s (with ABD +
MMS or manual dosing)
ABD transfer speed to furnace controlled
automatically via flame sensor or soft-
ware settings
0.2 to 2 µl/s
Recommended: 0.5 µl/s (with
Autoinjector)
Solids
Fixed/transfer speed to furnace con-
trolled automatically via flame sensor or
software settings.
Non-pressurized gases
1 to 40 ml/min
Recommended: 20 ml/min
Pressurized gases
Fixed (with GSS/LPG combi module)
1 to 40 ml/min
Recommended: 20 ml/min (with
GSS module and GSS adapterbox)
LPG
Solid
AOX
Fixed/transfer speed to furnace con-
trolled via software settings for the ABD
Measuring gas dryer
TS, TN, TC, EC/OC
Membrane dryer
TX, AOX, EOX
Concentrated sulfuric acid
TOC, NPOC, TIC
Condensation through Peltier cooling
Detection modules
Total nitrogen TN
N module 5100
Chemiluminescence
Total sulfur TS
S module 5100 basic and
S module 5100 MPO
UV fluorescence
S module 5100 coulometric Coulometric titration
AOX, EOX, TX, TCl, TOX, TIX Cl module 5100
Micro-coulometric end
point titration
(argentometry)
Total carbon TC, EC/OC
C module 5100
NDIR (non-dispersive
infrared absorption
spectrometry)
Total carbon TC, TIC, TOC,
NPOC, EC/OC
TOC module 5100
NDIR (non-dispersive
infrared absorption
spectrometry)
Sampling modules
Liquids
Multi Matrix Sampler
MMS
Auto scaling
Autoinjector
Semi-automatic
Solids
ABD + MMS
Auto scaling
Automatic Boat Drive
ABD
Semi-automatic
Non-pressurized gases
GSS module