multi EA 5100
Troubleshooting
103
Error
code
Error message/cause
Remedy
273
Auto-injector syringe: Syringe drawn incorrectly
(Autoinjector only)
Syringe not drawn back all the way
Draw syringe back all the way
Insert syringe
274
Autoinjector: No connection
Auto-injector coupling not found
Check connection
Acknowledge message
Initialize
No Autoinjector AI-EA was found
Autoinjector AI-EA not connected or faulty
Check connection
Acknowledge message
Initialize
275
Autoinjector: No syringe detected
Syringe not drawn
Syringe detection faulty
Repeat sampling with drawn syringe
Test a different syringe
300
Temperature controller: No connection
Communication error
Acknowledge message
Initialize
304
Temperature controller: Communication error
Temperature cannot be set
Acknowledge message
Initialize
400
Syringe pump: No connection
Communication error
Acknowledge message
Initialize
401
402
403
404
407
Syringe pump: Initialization
Syringe pump: Invalid command
Syringe pump: Invalid operand
Syringe pump: Invalid command sequence
Syringe pump: Device not initialized
Communication error
Syringe pump faulty
Acknowledge message
Search for cause and remedy fault
Initialize
409
Syringe pump: Pump sluggish
Gas hose clogged or kinked
Syringe pump faulty
Acknowledge message
search for cause and remedy fault
Initialize
410
Syringe pump: Valve sluggish
Syringe pump faulty
Valve faulty
Acknowledge message
Search for cause and remedy fault
Initialize