multi EA 5100
Troubleshooting
101
Error
code
Error message/cause
Remedy
201
Gas box: Fault when setting the target flow
Communication to gas box interrupted
Acknowledge message
Initialize
202
203
204
205
Gas box: Conversion fault 1
Gas box: Conversion fault 2
Gas box: Conversion fault 3
Gas box: Conversion fault 4
Communication faulty (readout of flows from gas
box faulty)
Acknowledge message
Initialize
206
Gas pressure error
Overpressure in the analysis system due to
blocked gas paths
WARNING! Extreme caution is required during
system overpressure! Never switch off a device
subject to overpressure! Otherwise, risk of injury
to operating personnel and damage to the analy-
sis system is present.
220
Sampler: no connection
Communication interrupted after autosampler de-
tection during initialization
Acknowledge message
Initialize
222
Boat: Broken
Boat defect during removal from the combustion
tube (only if boat sensor is used)
Remove the broken boat from the system
Initialize
223
Sampler: Incorrect peak value
No syringe inserted
Insert syringe in autosampler
Initialize
Dosed volume in the method greater than in-
serted syringe volume
Adjust dosed volume or insert suitable syringe
Initialize
A method for dosing liquids is to be activated and
the gripper and solids track are still inserted
Insert sample rack for liquid samples
Insert syringe
Initialize
224
Sampler: Incorrect gripper
No gripper inserted
Insert gripper in autosampler
Initialize
The method for solids is to be enabled and the sy-
ringe and sample rack for liquid samples are still
inserted
Insert solids rack
Insert gripper
Initialize
226
Sampler: Runtime exceeded
Completion message for the autosampler move-
ment takes too long (autosampler faulty)
Record log files
Inform customer service department