Maintenance and care
multi EA 5100
144
}
Switch off the detection module via the power switch.
}
Remove the left side panel. Remove the 4 screws to do this. Remove the protective
grounding cable and remove the side panel.
}
Remove the communication cable from the ozone generator.
}
Remove the 2 hoses from the ozone generator: "O
2
in" and "O
3
out". The "O
2
in" hose
connection is color-coded.
}
Remove both angled FAST connectors from the ozone generator.
}
Remove the screw the ozone generator is fastened to the base plate with (see ar-
row).
}
Carefully push the old ozone generator out of the module and remove it. Insert a new
ozone generator in the detection module. Mount the ozone generator in reverse or-
der. Replace the FAST connectors with new ones when doing so.
After replacement, check the system tightness:
}
Connect the detection module to the basic module.
}
Switch on both modules and allow them to run in for approx. 30 min.
}
Wet a strip of the indicator paper with distilled water and hold it to the ventilator on
the rear of the module for approx. 30 s.
}
Also check the gas outlet of the detection modules with the test strip.
}
If the strip turns blue, ozone is leaking from the module. If this is the case, switch off
the module, ventilate the room and check the proper connection of the hoses on the
ozone generator.
ü
The detection module is ready for operation again.
1
2
3
Figure 68Replacing the ozone generator
1 Oxygen connection (O
2
in)
2 Ozone outlet (O
3
out)
3 Communication cable (to the PCB)