Section 7: Foundation
Fieldbus
April 2017
Instruction Manual
IM-106-880C, OI
98
Mapping of
PWA
NV Memory
Failure
NV Writes
Deferred
High
Electronics
Temp-
erature
Transducer
Block:
EEPROM
Corrupt
Resource
Block:
1)
Manufacturing
2)
Block
Integrity Error
3)
NV Integrity
Error
4)
ROM
Integrity Error
Resource
Block:
1)
NV Writes
Deferred
Transducer
Block:
High Electronic
Temperature
FF/
Device
Alerts
FF/
Device
Alerts
Device
Alerts
The non-
volatile
parameter
storage on the
CPU board has
become
unreliable.
This alert
indicates that
the electronics
temperature
has exceeded
80°C. The
device will
cease to
operatereliably
beyond 85°C.
At startup, wait 2
minutes with
power applied and
then cycle power
again or reset
device.
Evaluate
mounting
location and
environment.
This alarm will generally occur during a
startup condition. Rarely, the device could be
powered down while a parameter is being
stored to the non-volatile memory. The
parameter will then be tagged as bad on the
next power on and the memory contents will
be written with default parameters.
Calibration data may be lost and the unit
should be recalibrated. If the unit does not
recover automatically, the memory may be
faulty and the electronics stack should be
replaced. Refer to Section 8, Overview page
165 for details.
A high number of writes has been detected
to non-volatile memory. To prevent
premature failure of the memory, the write
operations have been deferred. The data will
be saved on a 6 hour cycle. This condition
usually exists because a program has been
written that writes to Function block
parameters not normally expected to be
written to on a cyclic basis.
The device may require special mounting
considerations if installed in a very hot
location.
ADC Failure
Transducer
Block:
1)
ADC
Timeout Error
2)
ADC
Reference Error
Device
Alerts
This alert
indicates faulty
operation of
the device
electronics.
Cycle power or
replace the
electronics stack.
The Analog to Digital Converter (ADC) is
continuously monitored by the device for
correct operation. Refer to Section 8,
Overview page 165 for details.
Alerts
Alarms
AMS
Tab
What does
the alert
indicate?
Recommended
Action
Description
Calibration
Transducer
Calib-
This alert
Check instrument Oxygen cells will degrade over time due to
Recommended Block:
ration
indicates that
accuracy and/or
aging and corrosion. An increasing cell
Alerts
the sensor
calibrate.
resistance is a good indicator of reduced cell
Calibration
resistance has
performance. As the cell impedance
Recommende
changed by a
increases, the cell output falls off and
d
predetermined
response time increases. Calibrating the
amount since
instrument will compensate for the increased
the last
cell resistance. If using the device with an
calibration.
IMPS or SPS calibration sequencer, increased
cell impedance can automatically trigger a
calibration.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......