Section 5:
Calibration
April 2017
Instruction Manual
IM-106-880C, OI
78
COe D/A trim procedure using
HART
5.7.2
COe D/A trim procedure using HART
Use the following procedure to perform the COe D/A trim procedure using the Field
Communicator. If necessary, refer to
Section 6, Field Communicator
, Field Communicator,
for the HART menu tree.
NOTE:
To select a menu item, either use the up and down arrow keys to scroll to the menu item and
press the right arrow key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
1.
From the DIAG/SERVICE menu, select D/A TRIM. Press the up or down arrow to select COe D/A
Trim.
2.
Press the right arrow key to start the procedure. (If you wish to exit D/A Trim with no
changes, select ABORT.)
3.
The Field Communicator displays WARNING: Loop should be removed from automatic
control. Remove the Hazardous Area OCX 8800 from any automatic control loops to
avoid a potentially dangerous operating condition and press OK.
4.
The Field Communicator displays Connect reference meter to Combustibles output.
5.
Remove the electronics housing cover.
6.
Refer to
Figure 2-6
. Connect a digital multimeter to read the milliamp output from the
COe D/A converter circuit. Connect the positive lead to the AOUT2+ terminal and
connect the negative lead to the AOUT2- terminal. Then, press OK at the HART
communicator.
7.
The Field Communicator displays Setting Fld dev output to 4 mA. Press OK. Read the
COe milliamp output at the digital multimeter. Enter the reading at the Field
Communicator and press ENTER. (Select ABORT to exit without changes).
8.
The Field Communicator displays Setting Fld dev output to 20 mA. Press OK. Read the
COe milliamp output at the digital multimeter. Enter the reading at the Field
Communicator and press ENTER. (Select ABORT to exit without changes).
9.
The Field Communicator displays Setting Fld dev output to 4 mA. Press OK.
10.
The Field Communicator displays Fld dev output 4.00 mA equal to reference meter?
Using the up or down arrow, select 1 Yes or 2 No and Press ENTER. If No, the process
repeats from step 6.
11.
The Field Communicator displays Setting Fld dev output to 20 mA. Press OK.
12.
The Field Communicator displays Fld dev output 20.00 mA equal to reference meter?
Using the up or down arrow, select 1 Yes or 2 No and Press ENTER. If No, the process
repeats from step 7.
13.
The Field Communicator displays NOTE: Loop may be returned to automatic control.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......