Repair Electronics
Housing
April 2017
Instruction Manual
IM-106-880C, OI
21
4
Sensor Housing Leak
Test
9.4.15
Sensor Housing Leak Test
1.
Install 1/4 NPT cap on dilution air inlet fitting. Install a 1/4 NPT cap on sample tube (2,
) or plug 1/4 NPT sample inlet port. Capped or plugged ports must be air
tight.
2.
If not in place, install exhaust tube (3,
) in exhaust port according to the
instructions provided.
3.
Connect a calibrated manometer to the CAL GAS inlet port.
4.
Connect and apply clean instrument air at 45 psig (310 kPa gage) to the instrument air
inlet fitting.
5.
Observe the manometer reading. The reading should be from 10 to 13 inches, Water
Column. Locate and correct leaks if the reading is less than 10 inches WC.
9.5
Repair Electronics Housing
Use the following procedures to remove damaged components from the Hazardous Area OCX
8800 electronics housing and to install new replacement parts. Disassemble the unit only as
needed to replace damaged components. Use the assembly procedures that apply to install
replacement parts and reassemble the unit.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......