Instruction Manual
IM-106-880C, OI
Section 2:
Installation
April 2017
Electrical
Connections
25
To maintain explosion-proof protection of the OCX88C in hazardous areas, the sample
entering the sensor housing must not exceed 195
C (383
F).
NOTE:
To maintain proper earth grounding, ensure a positive connection exists between the sensor
housing, the electronics housing, and earth. The connecting ground wire must be 14 AWG
minimum. Refer to
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105 ºC (221 ºF).
2.4.1
Electrical Connections
Electrical connections, power and communications are made to the electronic enclosure. The
connections are made through two 3/4 NPT ports in the enclosure using fittings and cables
provided by the customer. Cable installation must meet NEC, IEC and/or other applicable
national or local codes for Class I, Zone 1, IIB +H2 T3/T6 permanently mounted equipment.
2.4.2
Connect Line Voltage
The Hazardous Area OCX 8800 operates on 100 to 240 VAC line voltage at 50 to 60 Hz. The
power supply requires no setup. Connect the line (L wire) to the L terminal, and the neutral (N
wire) to the N terminal on the AC power input terminal block in the electronics housing.
Connect the ground (G wire) to the ground stud in the electronics housing as shown in
2.4.3
Connect Output Signals
The Hazardous Area OCX 8800 may be provided with either two 4-20 mA signals with HART on
the O2 signal or a single Foundation Fieldbus signal. Connect the output terminals in the
electronics housing as shown in
. Use individual shielded twisted wire pairs. Terminate
the shield at the electronics housing.
2.4.4
O
2
4-20 mA Signal
One 4-20 mA signal represents the O
2
value. Superimposed on the O
2
signal is the HART
information accessible through a Model 375 Handheld Communicator or AMS Device Manager
software. The O
2
signal is at the AOUT 1 terminals.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......