Section 5:
Calibration
April 2017
Instruction Manual
IM-106-880C, OI
66
Operator - Initiated
Autocalibration
between automatic COe calibrations. To disable automatic calibration for O2 and COe,
enter 0 for both CalIntrvl parameters.
5.
If desired, the O2NxtCalTm and the COeNxtCalTm (next calibration time) parameters
can be changed to synchronize a calibration at a specific day or time.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that
O2 and COe are NOT calibrated simultaneously..
NOTE:
To select a menu item, either use the up and down arrow keys to scroll to the menu item and
press the right arrow key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
6.
From the O2 CALIB PARAMS screen, select CalIntrvl.
7.
At the prompt, input a time interval (in hours) at which an automatic O2calibration will
occur and press ENTER.
8.
From the DETAILED SETUP screen, select COE CALIB PARAMS.
9.
From the COE CALIB PARAMS menu, select CalIntrvl.
10.
At the prompt, input a time interval (in hours) at which an automatic COe calibration
will occur and press ENTER.
5.4
Operator - Initiated Autocalibration
An operator can initiate an automatic calibration at any time provided that the unit is equipped
with calibration solenoids.
5.4.1
Autocalibration using HART
To initiate an automatic calibration using HART/AMS, perform the following steps on the HART
menu tree. Refer to Section 4, Using HART Communications, for the HART menu tree.
1.
Select DIAG/SERVICE from DEVICE SETUP menu.
2.
Select CALIBRATE from the DIAG/SERVICE menu.
3.
Select PERFORM CAL from the CALIBRATE menu.
4.
Select CAL METHODS from the PERFORM CAL menu.
5.
From the CAL METHODS menu, select the type of calibration desired: O2 Calibration,
COe Calibration, or O2 and COe Calibration.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......