Instruction Manual
IM-106-880C, OI
Section 7: Foundation
Fieldbus
April 2017
Transducer Block
Parameters
10
7
O2_SLOPE
34.5-57.5
mV/Decade
The O
2
calibration slope.
O2_T90
0-300
Seconds
The amount of time that the O2 process variable will take to reach
90% of the actual process variable.
O2_TEMP_MAX
°C
The highest O
2
temperature read since power on.
O2_THERMOCOUPLE_
INPUT
mV
The raw value of the O
2
temperature input.
O2_TIME_TO_NEXT_CAL
0-9999
Hours
The time until the next automatic calibration of the O
2
sensor.
O2_TOL_CHECK
0: No – 1: Yes
O
2
calibration gas tolerance check.
OPERATING_MODE
Enumerated
Device Operating Mode. See
PCD_COUNTER
Power cycle drop counter.
PCDC_ENABLE
0: No – 1: Yes
Enable/disable power cycle drop detect.
PCN_COUNTER
Power cycle noise counter.
SB_HTR_DUTYCYCLE
Sample Block heater duty cycle.
SB_SETPOINT
°C
Sample Block heater temperature set point.
SB_TEMP
°C
The temperature of the sample line.
SB_TEMP_MAX
°C
The highest Sample Block temperature read since power on.
SB_THERMOCOUPLE_
INPUT
mV
The raw value of the sample line temperature input.
SENSOR_HOUSING_TEMP
°C
Sensor Housing temperature (Valid only with Type 3 sensor).
SENSOR_HOUSING_
TEMP_INPUT
SENSOR_HOUSING_
TEMP_MAX
mV
Sensor Housing CJC voltage (Valid only with Type 3 sensor).
°C
The highest Sensor Housing temperature read since power on.
SENSOR_HOUSING_TYPE 0, 1, 2, 3
Enumerated
This is the Sensor Housing type setting through the DIP switch.
(0=Type 1, 1=Type 2, 2=Type3, 3=Invalid)
SOLENOIDS_PRESENT
0: No – 1: Yes
Enumerated This determines whether a calibration cycle will automatically step
through, turning solenoids on and off to switch test gas, or wait for
an operator to manually switch gases and acknowledge.
ST_REV
See FF-891 section 5.3.
STATS_ATTEMPTS
Total number of messages sent to the transducer a/d board.
STATS_FAILURES
Total number of failed a/d board message attempts.
STATS_TIMEOUTS
Total number of timed out a/d board message attempts.
STRATEGY
See FF-891
section 5.3
.
TAG_DESC
See FF-891
section 5.3
.
TRANSDUCER_
DIRECTORY
See FF-903 section 3.3.
TRANSDUCER_TYPE
See FF-903
sections 3.3
.
UPDATE_EVT
See FF-891
section 5.3
.
VERIFY_STATE
0 through 5
Enumerated
The current state of the Calibration Check. (0=Idle 1, 1=Flow High
O
2
, 2=Flow Lo O
2
, 3=Flow High COe, 4=Purge Gas, 5=Done)
VERIFY_STATE_STEP
0, 1, 2,.3, 6
Enumerated Initiates a calibration verify of O
2
or Combustibles gas. (0=Start Flow
High O
2
, 1=Start Flow Lo O
2
, 3=Start Flow High COe, 3=Purge Gas,
6=No Effect)
VERIFY_STATE_TIME
Seconds
Time remain for the current Calibration Check state.
XD_ERROR
See FF-903
section 3.3
.
Parameter Mnemonic Valid Range Units
Description
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......