COe Purge/ Zero
Feature
53
3.9
COe Purge/ Zero Feature
This feature provides a way to periodically flood the COe sensor with air to perform two
functions; a) provide additional oxygen to help burn off any combustible residue from the COe
sensor, and b) allow for optional adjustment of the COe calibration constant. If the OCX8800 is
configured to update the COe calibration constant, only the constant is updated. The COe
calibration slope is not affected. To update both the constant and slope a full calibration must
take place.
The feature uses the calibration solenoid that is also used for high O2 test gas and COe zero gas.
For the feature to work properly, instrument air is used as the high O2 test gas. This also requires
the high O2 test gas value to be set at 20.95%.
As an option, a two way valve may be installed to switch the high O2 test gas between the
normal calibration gas and instrument air. This allows the Hazardous Area OCX 8800 to use a
specified calibration gas for calibration, then instrument air for the COe zero feature. Switching
between the two gases must be manually coordinated between scheduled calibrations and COe
zero events.
When the COe zero feature is used, special pneumatic connections are required. See
, for details.
The COe zero feature is only valid if the Hazardous Area OCX 8800 is supplied with calibration
solenoids and the solenoids have been activated.
A Hazardous Area OCX 8800 shipped from the factory must be configured before the COe zero
feature can be implemented. This same process must be performed any time a replacement
card stack is installed.
During the COe Zero Function the analog output signals may track the oxygen and
combustibles readings if configured to do so. To avoid a potentially dangerous operating
condition, the Hazardous Area OCX 8800 must be removed from the automatic
combustion control loop before performing the COe Zero Function procedure.
At the completion of the COe Zero Function, the COe analog output signal will change if
the Zero Update parameter is set to "Yes".
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......