configures the sample line heater control circuit. All switches are accessible through holes in the
electronics box.
Remove power from the Hazardous Area OCX 8800 before changing defaults. If defaults
are changed under power, damage to the electronics may occur.
Verify that the following switch settings are correct for your Hazardous Area OCX 8800
installation:
Figure 3-1 Hazardous Area OCX 8800 Defaults - HART Electronics
SW1 The two settings are internally or externally powering the O
2
4-20 mA signal. The factory
setting is for the O
2
4-20 mA signal to be internally powered.
SW2 The two settings are internally or externally powering the COe 4-20 mA signal. The factory
setting is for the COe 4-20 mA signal to be internally powered.
SW3 The factory sets this switch as follows:
•
Position 1 determines the O
2
4-20 mA signal rail limit. The settings are high, 21.1
44
mA, or low, 3.5 mA. The factory setting is low, 3.5
Ve
m
rif
A
y
.
Configuration - HART
Electronics
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......