Instruction Manual
IM-106-880C, OI
Repair Sensor
Housing
April 2017
Install COe Sensor
Assembly
20
9
Figure 9-20 COe Sensor Holder Alignment
8.
Install and fasten thermocouple (8,
9.
Position band heater as shown in
and tighten band
heater clamp screw. The heater insulator (9) end joint must line up with the band gap
of the COe band heater (10).
10.
Reconnect the COe sensor, thermocouple, and heater wires at the sensor housing
terminal blocks. Refer to
11.
Install and fasten the COe insulator (1,
) around COe sensor assembly
(5). All wiring must remain outside of the insulator.
12.
If terminal block mounting (13,
) was moved, reinstall with two
base mounting screws.
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......