Section 1: Description and
Specifications
April 2017
Instruction Manual
IM-106-880C, OI
14
Specification
s
Sensors Housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to clean, dry atmosphere, two 3/4-14 NPT
conduit ports
Electronics Housing
Type 4X, IP66 with fitting and pipe on reference exhaust port to clean, dry atmosphere, two 3/4-14
NPT conduit ports
Certifications
Complies with the following standards:
•
22.2 No. 94-M91; 22.2 No. 60529:05; ANSI/ISA S82.02.01; ANSI/ISA 12.00.01:2002; ANSI/ISA
12.22.01:2002;UL No. 50 (Ed 10); CAN/CSA C22.2 No. 61010-1-12; CAN/CSA C22.2 No.
60079-0:15; CAN/CSA C22.2 No. 60079-1:16; ISA 60079-0 (12.00.01)
– 2013; ISA 60079-0
(12.00.01)
– 2013; ISA 60079-1 (12.22.01) – 2009 (R2013) ISA 61010-1-12
•
ATEX: EN60079-0:2012; EN60079-1:2007
•
IECEx: IEC 60079-0: 2004, Edition 4; IEC 60079-1: 2014, Edition 7.0
•
FM: Class 3600:1998; Class 3810:2005; ANSI/ISA 12.00.01:2005; ANSI/ISA 12.22.01:2002; ANSI/
NEMA 250:1991; ANSI/ISA 60529:2004
Electrical Noise
EN 61326-1, Class A
Line Voltage
Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or jumpers required, 3/4-14 NPT conduit
port Pollution Degree
2
Over Voltage Category
II
Relative Humidity
5 to 95% (non-
condensing) Isolated Output
Oxygen
4-20 mAdc, 950 ohm maximum, with HART or Foundation Fieldbus capability
only Combustibles
4-20 mAdc, 950 ohm maximum (not present with Foundation Fieldbus)
Alarm
Alarm output relay - dry contact, form C, 30mA, 30VDC
capacity Power Consumption
750 W maximum
1.
All static performance characteristics are with operating variables constant. Specifications subject to change without
notice.
Specifications
1
Blowback Air (optional)
Clean, dry, instrument quality air (20.95% O2) regulated to
≥
60 psi (413 kPa) or greater and ambient
temperature of
≥
0 °F (-18 °C)
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......