Section 4: Using the
LOI
April 2017
Instruction Manual
IM-106-880C, OI
60
LOI Status
Codes
4.1.5
LOI Status Codes
The LOI display shows a status code in the lower right hand corner of the display. There are nine status codes to
indicate the existing status of the device during operation. The status code descriptions are shown in
Table 4-1. LOI Status Codes
AL
Alarm
- The device is in a recoverable alarm state.
BL
Blowback
- A blowback cycle is active.
CA
Calibration
- A calibration cycle is active.
CV
Calibration Verify
- A calibration verify task is in progress.
NM
Normal
- The device is in a normal operating mode.
PO
Power On
- A system level initialization sequence is active. This will continue for
several seconds.
SF
System Fault
- The device is in a non-recoverable alarm condition. The unit
must be reset or power must be cycled off and on to resume operation.
ST
Stabilize
- The device heater control is stabilizing (after warm up). Sensors are
warming up to operating temperature.
WU
Warm Up
- The device heaters are ramping up to operating temperature.
CZ
COe Zero
- The COe Zero cycle is active.
4.2
LOI Menu Tree
This section consists of a menu tree for the LOI on the Hazardous Area OCX 8800,
. This menu is specific
to the Hazardous Area OCX 8800.
4.2.1
First Column Submenus
From the operating display (O
2
% and COe ppm), the left-pointing Enter key is the only option to move into the first
column submenus of the LOI menu tree. The first column contains three submenus: SENSOR DATA,
sheet 1 of 4, CALIBRATION, sheet 2 of 4, and SYSTEM, sheets 3 and 4 of 4. From the operating display, SENSOR
DATA is displayed when the right-pointing key is selected. Use the up or down-pointing key to move to the other
first column submenus.
4.2.2
Second Column Submenus
From the first column submenus, selecting the right-pointing key moves the display into the second column
submenus. The up and down-pointing keys allow the display to move to the second column submenus of the first
column submenu selected. The left-pointing key moves the display back to the first column submenu.
CODE
DESCRIPTION
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......