Section 3: Configuration and
Startup
April 2017
Instruction Manual
IM-106-880C, OI
3.8
Calibration Tolerance Feature
The calibration tolerance feature provides a mechanism to fail a calibration if the calibration
measurement does not fall within a specific tolerance of the test gas value. The tolerance is
preset within the Hazardous Area OCX 8800 software and is not user adjustable. The tolerance
is different between the oxygen and combustibles test gases. For oxygen, the calibration will fail
if the measured value differs by more than ±10% of the configured value. For combustibles, the
calibration will fail if the measured value differs by more than ±30% of the configured value.
A Hazardous Area OCX 8800 shipped from the factory must be configured before the
calibration tolerance feature can be implemented. This same process must be performed any
time a replacement card stack is installed.
3.8.1
Configuring the Calibration Tolerance Feature with the Field
Communicator - HART
1.
Use the 375 Field Communicator or AMS software to access the HART menu.
2.
From the DETAILED SETUP menu, select CAL SETUP.
3.
From the CAL SETUP menu, select O2 CAL PARAMS.
4.
To enable the calibration tolerance feature for the oxygen calibration, from the O2 CAL
PARAMS, select O2 Tol Check. Select On to enable the calibration tolerance feature.
5.
Back out to the CAL SETUP menu, select COe CAL PARAMS.
6.
To enable the calibration tolerance feature for the combustibles calibration, from the COe
CAL PARAMS, select COe Tol Check. Select On to enable the calibration tolerance feature.
3.8.2
Configuring the Calibration Tolerance Feature with the Field
Communicator - Fieldbus
1.
Use the 375 Field Communicator or AMS software to access the Fieldbus menu.
2.
From the TRANSDUCER block menu, select O2 CAL/COe CAL.
3.
From the O2 CAL/COe CAL menu, select O2 CAL SETUP/COe CAL SETUP.
4.
From the O2 CAL SETUP/COe CAL SETUP menu, select O2 Tolerance Check/COe
Tolerance Check. Select Yes to enable the calibration tolerance feature.
3.8.3
Configuring the Calibration Tolerance Feature with the LOI
1.
Use the "Z" pattern to enter the LOI menu tree.
2.
From the SYSTEM menu, select Calib Setup.
3.
From the Calib Setup menu, select the following:
•
O2 Tol Check - Select Yes to enable the calibration tolerance feature for the
oxygen calibration.
•
Comb Tol Check - Select Yes to enable the calibration tolerance feature for the
combustibles calibration.
52
Calibration Tolerance Feature
Summary of Contents for Rosemount OCX 8800
Page 2: ......
Page 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Page 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Page 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Page 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Page 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Page 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Page 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Page 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Page 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Page 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Page 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Page 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Page 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Page 305: ......