TRANSIENT3000
55/162
5.4.1 TRA3000 not equipped with EXT-TRA3000 D:
No DIP DC (Interrupt) will be visible in the display. The d.c. supply of the EUT can be turned ON/OFF with
the PWR button on the front of the TRA3000. PWR2 is out off order for d.c. supply.
When on + input a negative voltage of the d.c. supply is connected an error message will occur as shown
below.
Without any guarantee from EMCP of correct operation the lower voltage limit of 20V can be extended by
selcting “No” on Power 1 see display below.
5.4.2 TRA3000 equipped with EXT-TRA3000 D
The display indicates the different possibilities of DIPS and Interruptions
5.4.3 DIP DC (Interrupt) without PS3 power supply source of EMC PARTNER
On his own risk the customer can use his own power supply.
In DC Mode only short Int. can be selected. (
Volt dips
and
Volt var.
are only selectable when a power
supply PS3 is connected to TRA3000)
Renmark: High-Z at 0% must be selected
off
For voltages lower < 20V No in Power 1 and Power2 must be activated. For power >20V Mains must be
selcted.
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...