TRANSIENT3000
80/162
6.2.2.8 Editing DIPS longer than one period
Long DIPS are interruptions with a duration longer than one period. The duration is specified in ms. The
turn ON and OFF angles can be selected.
loaded test
EUT power v, i
test parameters
tests parameters
-400.00
-300.00
-200.00
-100.00
0.00
100.00
200.00
300.00
400.00
Duration is the length of time of the lowered voltage.
Repetition defines how often the disturbance occurs.
6.2.2.5.1 Definition duration, repetition, test time
When the soft key F6 has been pressed:
loaded test
EUT power v, i
test parameters
tests parameters
6.2.2.5.2 DIP Begin, DIP End
The TRA3000 F-S-D-V-C processor calculates the number of disturbed periods as a function of the chosen
duration. The condition for a correct test is the correctly selected power frequency, in the range of (16
2
/
3
to
Begin
End
DIP in ms
Test time
Duration
Repetition