TRANSIENT3000
25/162
1.3.9 Common mode test with EXT-TRA3000 C
Caution
To avoid any damage of the EUT during CM test set-up the EUT power cord on the rear side of the
TRA3000 shall be removed. When the CM test is selected the power line is disconnected and the PWR1
and PWR2 can not switched “ON”. As soon as an other test e.g. Surge is selected the PWR1 and PWR2
can be activated and the power voltage is on L at the front of the TRA3000.
1.3.9.1 General Generator Specification
Voltage setting range
0.1 to 35V
0.1 V
resolution
Source impedance
50 Ohm
± 10%
Synch turn ON/OFF
0°
± 5%
of test voltage
Operation mode: Continuous
Leves V
1 up to 4
Open circuit output voltage range
Vmin 0.1V
Vmax 30V
-10%
+10%
Test frequencies
DC, 16,7Hz, 50Hz and
60Hz
Test durtion
1 up to 30’000s
Test time
1 up to 30’000s
Operation mode: Short
Leves V
1 and 2
Test frequencies
DC, 16,7Hz, 50Hz and
60Hz
Test duration
1 up to 10s
Repetition
Du 1 s
Test time
1 up to 10s
Operation mode: fix frequencies
Range
DC up to 150kHz
1Hz
resolution
Operation mode: sweep
frequencies
See chapter below
Levels
1 up to 4
1.3.9.2 Characteristics and performance of the generator for d.c. tests
Test levels for continuous operation
1, 3, 10, 30 V
Test levels for short time operation
10, 30 V
1 to 10s
Selectable short duration
time
Switching time at d.c. ON/OFF
Between 1 - 5 µs
1Hz
resolution
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...