TRANSIENT3000
26/162
1.3.9.3 Characteristics and performance of the generator for a.c. tests
Test levels for continuous operation
1, 3, 10, 30 V
Test levels for short time operation
10, 30 V
1 to 10s
Selectable short duration
time
Test frequencies
16,7Hz, 50Hz and
60Hz
1.3.9.4 Characteristics and performance of the generator for tests in the frequency range 15 Hz-
150 kHz
Frequency range
15Hz up to
150kHz
-10%,
+10%
Test levels
1, 2, 3, 4,
Decade time
10s up to 1000s
Step Sizes of the preceding
frequency value
2 to 10%
Voltage changes
20dB / decade.
20dB / decade.
Decreases
Increases
15 up to 150Hz
1.5kHz up to 15kHz
No test level is defined below 15Hz, excluding d.c.
Generator output is short circuit protected
For further information about short duration disturbance test see EXT-TRA3000 C-SHORT instruction
sheet.
1.3.10 Power line limits on EUT power input of TRA3000
Power line voltages greater than 280V and frequencies higher than 70Hz are the limit for the power
dissipation of the SURGE circuit. The TRA3000 measures the voltage and the frequency. When higher
voltages or frequencies are measured the following modes will be activated:
Gating Mode
Gating Time
Gating activated
0
No Gating
F < 70Hz
& V < 280Vrms
1
0.5 Seconds
F < 70Hz
& V >= 280Vrms
2
0.1 Seconds
F >= 70Hz
Explanation of the gating mode:
The coupling path between the SURGE circuit and the power line will be opened and only closed during
Surge release.
Possible states during “RUN” modes:
0
Æ
1
0
Æ
2
1
Æ
2
1.3.11 Synchronisation of TRA3000 to mains frequencies
Synchronisation to EUT power input
A minimum voltage of 30V rms must be applied to the power input on the rear of the TRA3000 (PWR1 or
PWR2) to synchronize the Surge to the mains. The LED within the connector on the front panel indicates a
proper synchronisation.
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...