6.2.2.6 Synchronisation
Output
loaded test
EUT power v, i
test parameters
tests parameters
Note: The 0° is defined as the angle at which the Surge is superimposed on the power supply depending
on coupling path selected in the 3P CDN.
Selection of coupling path
loaded test
EUT power v, i
test parameters
tests parameters
When the 2 x V-peak are selected "on" the voltage of the L
Æ
N voltage will be doubled as specified in the
standard for common and differential tests.
6.2.2.3.3 Coupling path definition SURGE
When several coupling paths are activated „ON“, they will be chosen step by step. The sequence
corresponds with the sequence listed in the „Main“ menu. When Impulse -Out = On, the impulses stay at
the high voltage (banana plugs). The coupling kit or the three phase filter can be connected on the high
voltage SURGE output.
At Power Synchro =ON, the surges are released synchronous to the main frequency.
At SURGE Trigger = Manual, the surge is not
automatically released after the repetition time, but
by operation of the Man-Trigger button.
6.2.2.3.3 Definition synchronisation
t
Synchro freq 1/T
Synchro angle
L
N
L - N
N - PE
PE
L - PE
Summary of Contents for EXTTRA3000 E
Page 8: ...TRANSIENT3000 8 162...
Page 102: ...TRANSIENT3000 102 162...
Page 106: ...TRANSIENT3000 106 162...
Page 116: ...TRANSIENT 2000 116 162...
Page 118: ...TRANSIENT 2000 118 162...
Page 142: ...11BRemote Ports 142 162 All Generators of EMCP can be loaded and demonstrated...
Page 148: ...12BAppendix and Corrections 148 162 13 1 6 VARIATION Specification IEC 61000 4 11 Ed 2...
Page 150: ...12BAppendix and Corrections 150 162...
Page 156: ...14BIndex 156 162...
Page 158: ...14BIndex 158 162...
Page 160: ...14BIndex 160 162...
Page 162: ...14BIndex 162 162...