TRANSIENT3000
87/162
-400.00
-300.00
-200.00
-100.00
0.00
100.00
200.00
300.00
400.00
Smallest step is 1/f. (f=Mains Power frequency EUT)
The Begin and End angles remain constant as
selected in „Main menu“.
6.2.3.5.2 Definition duration ramp
The chosen DIP Begin and DIP End angles remain constant during a duration ramp. As a consequence,
the duration does not change continuously, but over one period.
Ramp: Ramp angle at DIP Begin:
loaded test
EUT power v, i
Duration ramp at DIP
At DIP Begin, the angle changes within the range specified.
-400.00
-300.00
-200.00
-100.00
0.00
100.00
200.00
300.00
400.00
6.2.3.5.3 Definition ramp
Ramp: Ramp angle at DIP End:
At End Ramp, the angle will be changed in steps of 5° from 90° to 0° when the EUT power is turned on.
With this mode the inrush current after a interruption can be investigated.
1 dip
1 dip
Repetition
Repetition
Dur. start
Duration
Dur. start+step
Begin+10