Functional description
5
gas-lab
Q1
Page 45
After confirmation the calibration with the 3
rd
calibration gas (L1-8K or a binary
gas mixture) proceeds automatically. The remaining time is indicated on the
displayed.
In any case “L1-8K” is shown on the display, even if you use a binary
gas mixture instead.
The adjustment starts after approx. 2 minutes for purging and lasts for 15
minutes, whereas the remaining time is displayed
The end of calibration is indicated by the following display:
Then the display changes to the following condition (same as before starting):
After the basic calibration is finished, the sensor system automatically switches
back to the process gas and starts measuring.
Now turn close the L1-8K or binary gas mixture cylinder and remove the tube on
both sides.
(Caution: Unpressurise the line first!)