Annex
11
gas-lab
Q1
Page 141
H 2005 T min
The hint is generated along with alarm A606 when the lower temperature alarm limit is
violated.
H 2006 T max
The hint is generated along with alarm A606 when the upper temperature alarm limit is
violated.
H 2007 UTCS min
The hint appears when the signal voltage of the thermal conductivity detector, violates the
low alarm limit. It is generated together with alarm A605 (data error), to discribe the exact
cause of the fault.
H 2008 UTCS max
The hint appears when the signal voltage of the thermal conductivity detector violates the
upper alarm limit. It is generated together with alarm A605 (data error), to mark the exact
cause of the fault.
H 2009 ICHref min
The hint appears when the signal voltage of the infrared sensor for hydrocarbons,
reference channel, violates the lower alarm limit. It is generated together with alarm A605
(data error), to mark the exact cause of the fault.
H 2010 ICHref max
The message appears when the signal voltage of the infrared sensor for hydrocarbons,
reference channel violates the upper alarm limit. It is generated together with alarm A605
(data error), to mark the exact cause of the fault.
H 2011 ICHmeas min
The hint appears when the signal voltage of the infrared sensor for hydrocarbons,
measurement channel, violates the lower alarm limit. It is generated together with alarm
A605 (data error), to mark the exact cause of the fault.