gas-lab
Q1
H 2012 ICHmeas max
The hint appears when the signal voltage of the infrared sensor for hydrocarbons, channel
violates the upper alarm limit. It is generated together with alarm A605 (data error), to
mark the exact cause of the fault.
H 2013 ICO2ref min
The hint appears when the signal voltage of the infrared sensor for CO2, reference
channel, violates the lower alarm limit. It is generated together with alarm A605 (data
error), to mark the exact cause of the fault.
H 2014 ICO2ref max
The hint appears when the signal voltage of the infrared sensor for CO2, reference
channel violates the upper alarm limit. It is generated together with alarm A605 (data
error), to mark the exact cause of the fault.
H 2015 ICO2meas min
The hint appears when the signal voltage of the infrared sensor for CO2 measurement
channel violates the lower alarm limit. It is generated together with alarm A605 (data
error), to mark the exact cause of the fault.
H 2016 ICO2meas max
The hint appears when the signal voltage of the infrared sensor for CO2 measurement
channel violates the upper alarm limit. It is generated together with alarm A605 (data
error), to mark the exact cause of the fault.
H 2017 Numeric ERROR
The message appears when the calculation of the target values exceeds error limits.
H 2018 Result ERROR
The hint appears when the calibration is wrong or missing. The calculation of the target
values (Hs, Rhon, CO2) is not possible. During measurement operating gas or revision
gas and it can be calculated not a virtual analysis. Simultaneously A605 (data error) is
generated.