9
Maintenance
Page 126
gas-lab
Q1
Proceed as described below:
A battery of the following type is required: lithium 3V CR ½ AA.
Save the device parameterisation by reading it out with GAS-WORKS, to be
on the safe side. Read out the archives, too.
Disconnect the supply voltage.
Loosen the 4 fixing screws on the back of the device. Two of these screws
could be hidden and secuted by (official) adhesive labels or seals.
Slightly withdraw the mounting rail.
Attention:
The boards are sensitive!
Avoid any contact with other components!
The battery compartment is down on the left of the board that is located on
the far right side (seen from behind). Detach the cover strap of the battery
compartment with a screwdriver. Remove the old battery. Now you have 15
minutes for the battery exchange, during which a data backup is
guaranteed by a capacitor. Insert the new battery.
Note:
When inserting the battery, pay attention to the correct polarity: The
positive pole (+) must point downwards! A wrong polarity will not be noticed
at first, but during the next power failure of more than 15 minutes the device
data will be lost.
Re-assemble the device and connect the supply voltage.