Operation
3
gas-lab
Q1
Page 13
The sequence of the LED status in the above table corresponds to the
sequence the error management keeps to: The system always indicates the
error with the highest priority. A pending error always takes precedence over an
error that is no longer relevant.
The exact meaning of the terms
alarm
,
warning
and
hint
is explained in Chapter
The calibration switch is on the lower right side of the front panel.
All
gas-net
devices are furnished with a two-level safety concept: All parameters
being protected by the calibration switch can only be changed if the calibration
switch is open. Such parameters are always modified with a PC or laptop and
the associated parameterisation software GW-GNET+.
Open the calibration switch by turning it anticlockwise as far as it will go. This
first level of the safety concept is important for devices used for legal metrology
and custody transfer. In this case, a seal may officially secure the calibration
switch.
The basic display of the device will automatically be invoked when you close the
calibration switch.
Note:
The
User lock
as the safety concept’s second level consists of one
numerical lock for each of the two contract parties. The user lock is, in contrast
to the calibration switch, implemented via the device software. This means that
the locks are defined via the device parameterisation and opened or closed via
the operator panel. Open locks allow the user to access certain parameters or
actions. All parameters being subject to the user locks can be changed when
both locks are or the calibration switch is open. .