The parameterisation of the evaluation computer defines which gas has to be
connected to which inlet. The process gas is assigned to path 1 and the internal
calibration gas (ultra-pure methane) to path 2. Nitrogen and further calibration
gases are injected via path 3 for performing the basic calibration. The test gas
can either be injected via path 1 or 3.
gas-lab Q1
exhaust gas
connection of
safety relief valve
power supply
connection
exhaust gas
connection of
gas-lab Q1
pressure controller
with integrated
safety relief and
safety shut-off valve
data line
connection
pressure gauge of
inlet pressure for gas-lab Q1
calibration gas 1
connection
process gas
connection
calibration gas 2
connection
release button of
safety shut-off valve
ex-junction box
Not only the sensor housing is installed on the mounting plate, but also
regulators for the inlet pressure of the different gases to be injected, safety shut-
off and safety relief valves as well as the EEx-e junction box for the
interconnecting cables (voltage supply, data link to the evaluation computer).