A607 Pressure failure
The alarm is provided for monitoring the pressure of the sensor system. His appearance
during violation oft the configuted limit values (upper / lower alarm limit) is set by the
parameter "Pressure error is warning". If this parameter is set to
“No”, the alarm is allways
generated (after a timeout), if no gas pressure is present in the instrument. When
"Pressure error is warning" = Yes and calibration gas is switched on, this alarm is replaces
by warning W2030 (see below). Such a pressure error with process gas connected
alwaysleads to the alarm A607 Pressure failure.
A647 GQ disturbed
The alarm indicates a general problem in the gas quality measurement. It is generated in
the following situations:
-
Any sensor raw value (eg ICHmeas, ICHref, ...) from the violates the configured
alarm limits.
-
A triggered 1-point calibration with methane has been canceled without success
(vioating a stability timeout).
A 670 CGM disturbed
The alarm indicates that the calibration gas measurement is disturbed. Examples:
-
Incorrect associated measuring equipment or incorrect serial number,
(in this case, H20021 is generated simultaneously)
-
Calibration time out (in this case W 753 is generated additionally)
H423 Interface monitoring
If this hint is pending , the device software can not recognize an interface or internal
board. This occurs, for example, if a card is broken, a connector is not plugged in or the
communication between the controller and sensorsystem is interrupted.A correct device
operation is no longer guaranteed.
H671 Pmax p-regulator
This hint (group signal) is generated when the safety relief valve (SBV) of the high-,
medium- or low-pressure reduction is triggered or it indicates t violation of the upper limit
of input pressure for the high-pressure reduction. Corresponding contact pressure meter
provided.