Technical data
10
gas-lab
Q1
Page 129
10 Technical data: Q1
10.1 Device type
The
gas-lab
Q1 measuring system is a flameless measuring instrument for
determining the quality of natural gases using an infrared absorption-based
measuring method. It consists of a heavy-duty housing, in which the actual
sensor unit is accommodated, and of an evaluation computer, which must be
operated in a non-hazardous environment. The system operates continuously
and fast and can be used for billing or controlling applications.
10.2 Sensor technology details (
gas-lab
Q1)
Measurements
17
Primary measurements: Heating value superior Hs =
gross calorific value GCV: 30 to 47MJ/m
3
Standard density
s :
0.71 to 0.97 kg/m
3
xCO
2
content: 0 to 5 % (optional 0 to 20 %)
Derived measurements:
Heating value inferior Hi= net calorific value NCV,
Wobbe number, methane number and calculated gas
analysis (10 components)
Suitable gases
2
nd
gas family (natural gases) according to G260
CH
4
:
75%
to
100%
C
2
+:
0%
to
15%
N
2
:
0%
to
20%
CO
2
:
0%
to
5% (Optional 0
– 20 %)
O
2
:
0%
to
2.0%
Others parts
<
0.1 %
17