Technical data
10
gas-lab
Q1
Page 133
Interfaces
DSfG interface with a maximum transmission rate of 115200 baud. Gas quality
data and data logging archives retrievable via DSfG.
DSS data interface for the connection to the COM interface of a PC or laptop
(for parameterisation when commissioning and for archive polling).
COM2 interface (serial interface to RS232C) for the connection of a modem with
a maximum transfer rate to V.34bis (33600 baud) and data compression to V.42
or NMP5.
DCF77 interface for the connection of a radio clock receiver.
Data logging
Integrated data logging function for logging billing and operating data.
Parameterisation
Commissioning and parameterisation are implemented via the GAS-WORKS
PC software. Parameter data records can be stored, documented and managed
in the PC under GAS-WORKS.