Functional description
5
gas-lab
Q1
Page 95
5.6.2.4
Display and operation
The
(Host) Communication
menu item for the
Data exchange
module is
intended for trouble shooting by experts and thus not explained in detail here.
Please get into contact with FLOW COMP if you need support or more detailed
information.
The
Protocol channel
submenu provides status information for individual
protocol channels via which a host is connected. Select the protocol channel to
be indicated via the
Protocol selection
list.
The subordinate menu item
Host
registers
gives detailed information on the data exchange via this protocol
channel, broken down into host registers. Either scroll through the individual
register positions using the arrow keys or select a particular start register for the
display via the
Register selection
submenu.
The
Error statistics
subordinate menu shows the number and type of the
communications errors that have been detected in case of the selected protocol
channel. You may reset the error statistics manually via the operator panel
(
Clear statistics
).