Functional description
5
gas-lab
Q1
Page 41
2. Zero point adjustment with nitrogen
While being in the gas quality main display press the menu key and activate the
Calibration
command in the invoked menu.
A display with measurements of the different sensors appears.
By pressing the menu key again, the following submenu appears.
Select the menu item 3
P-cal. start
and confirm.
Now the actual status of the 3-point calibration is indicated on the display. Here
e. g.: The required operation
–
switch on gas
(nitrogen channel 3)