8. Phased Array (PA) Mode
(optional – for D 1000-PA only)
8.1 Initializing PA Mode
In order to switch ON the PA (Phased
Array) mode, depress the
key and then
depress the right (grey) control key next to
the “
Phased Array OFF
” on the display.
Depending on the connected probe, select
the appropriate
Phased Array
mode (
16x1.0mm
/ 16x3.5mm / 32x1.0mm / 8/16x1mm / 16/32x1mm
).
8/16x1mm probe is equivalent do MWB 9x8.
In order to set default settings, depress
SETUPS
→
DEFAULT
and confirm
YES
.
If the PA probe is equipped with wedge,
depress the right control key (grey) next to
the “
PA WEDGE OFF
” on the display.
For 0 degree wedge, depress the left control
key (yellow). The following values will be set
automatically:
PA Z-OFFSETO=20mm; PA WF=10mm –
see 8.2.2 and 8.2.3.
Leave the “
PA MODE Chan ALL
” – this is
default mode with all 16 channels operated
as transmitter/receiver, or choose the appropriate combination of elements
The “
PA ACG
” (Phased Array Angle Corrected Gain) may be set to “
OFF
” or
“
DEFAULT
”. In the latter case, the gain will be automatically corrected depending on
the ultrasound beam angle in the dynamic mode.
NOTE: In order to correct function of S-Scan (Sector scan) the BSCAN MODE
must not be set to THICKNESS in
SETTING
>
DISPLAY
menu.
?