4.7 DISPLAY Sub-Menu (SETTING)
This sub-menu allows to set some basic
functions of the display.
DISPLAY Sub-Menu Items on the Screen
– Left Side:
4.7.1
LANGUAGE
Setting
of
interface
language
using
appropriate left or right button:
-
0 = English,
-
1 = Japan,
-
2 = Czech,
-
3 = Russian,
-
4 = Turkish,
-
5 = Chinese,
-
6 = German.
NOTES: A number of available languages depends on the localization made by
regional dealer.
The number and order of languages depends on custom setting and is
determined in the “lang_cfg.txt” file in the “/D1000/lang” folder. See also
10.5. Firmware Update.
4.7.2
UNITS
Setting of measuring units for time base and all calculated values – mm / inches /
microseconds (
µ
s). Available settings are as follows:
-
mm &
µ
s – push once the left (yellow) button;
-
mm
– initial (default) setting;
-
inch
– push once the right (grey) button;
-
inch &
µ
s – push twice right (grey) button.
NOTES: With the “
µ
s” option activated, the following data will be calculated and
displayed in microseconds: S-axis (sound path), positions of all gates,
positions of measured echoes, distance between echoes (measured
thickness), attenuation correction in dB/
µ
s. All other values remain
unchanged, e.g. for angle probe, P (projection distance) and D (depth) in
mm (inches), thickness diagram and the BSCAN mapping min/max.
When A.CALIB had been activated, the microsecond units will be
automatically switched off.
The time base settings in “
µ
s” a saved in setups separately, as well as units
mm/inch; setups saved with older SW versions (using Velocity 2000 m/s)
will be loaded with units set to “
µ
s” automatically.