Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
45/102
4.5.6
BSCAN data records
Each time after BSCAN had been activated, measured thickness values are recorded
to the memory in a temporary file.
In order to save actually measured data, depress the Display Freeze key
and save
the picture – see 3.6.2. Data of the last measured scan will be saved to the file with
the same name, as actual picture (Pnn.dddddd, but with “.txt” extension). The data
file length is limited to 500 kB.
Each line of the data file (.txt) contains measured value in mm (2 decimal positions)
or inches (3 decimal positions), depending on actual settings:
XXX.XX <cr><lf>
If incremental encoder (IRC) had been connected, then each line of the data file (.txt)
contains actual co-ordinate as well:
NNNN; XXX.XX <cr><lf>
The data file format enables simple import to MS Excel etc. Dot (not comma) is used
as a decimal separator.
For saving data to USB memory stick – see 10.3. Save Pictures (Screenshots) And
Data To USB Memory Stick.