Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
57/102
5.1.3
Functions of GATE 1 with Angle Probes
When the ANGLE function is “ON” and X-VALUE had been set in the Settings menu,
some values are displayed using corrected metrics:
-
(1) – S=XXmm – non-corrected sound
path – position of maximum echo peak
(marked with blue cross “
×
”) within the
GATE 1. The skip-leg position of the
echo peak is indicated next to the
value by \ / symbol. The position of
echo peak on the skip diagram is
indicated by small grey square, which
changes to red, when echo crosses
the gate level;
-
(2) – 1.P=XXmm – position and height of first echo peak (marked with white cross)
over the GATE 1;
-
(3) – D=XXmm – calculated depth of the indication (maximum echo peak within
the GATE 1, marked with blue cross “
×
”);
-
(4) – maximum echo peak (marked with blue cross “
×
”) relative amplitude against
the GATE 1 [dB] (not active in RF display mode);
-
(5) – maximum echo peak (marked with blue cross “
×
”) absolute height within the
GATE 1 [% FSH];
-
(6) – P – projection distance from the probe X-point to the maximum echo peak
(marked with blue cross “
×
”) within the GATE 1.
(6)
(5)
(4)
(1)
(2)
(3)