Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
52/102
4.8 CURVES Sub-Menu (SETTING)
This sub-menu allows settings of different
curves modes and parameters.
CURVES Sub-Menu Items on the Screen
– Left Side:
4.8.1
DGS
Switching ON / OFF the DGS function – see
7.4. DGS Function.
4.8.2
TCG
Switching ENABLE / DISABLE the TCG (Time Corrected Gain) mode – see 7.3.7.
TCG Modes (Time Corrected Gain)
4.8.3
REF TYPE
Setting the reflector type in the reference (calibration) block:
-
BW-K1 - back wall reflector R100 on the K1 (V1) standard block acc. to
EN 12223,
-
BW
- back wall reflector,
-
DSR
- disc shape reflector,
-
SDH
- side drilled hole.
4.8.4
REF SIZE
Setting the reference reflector size – for DSR and SDH only.
If the reference reflector size (set in this item) is greater than the probe equivalent
diameter D
eff
(see 4.9. PROBE Sub-Menu (SETTING), then the number white colour
changes to the red.
4.8.5
ERS DISP
Switching ENABLE / DISABLE the Equivalent Reflector Size (ERS) Display in the
Status window with active
DAC
(see 7.3. Distance-Amplitude Correction
Function (DAC)) or
DAC+TCG
(see 7.3.7. TCG Modes (Time Corrected Gain))
function.
The equivalent reflector size is displayed as
DSR
= XX.X
or
SDH
= XX.X
[mm], depending
on the reflector type set in the REF TYPE – see 4.8.3 above.