Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto,
© DRC Srl 2012
37/102
GATES Menu Items on the Screen – Right Side:
4.4.6
GAIN
Adjustment of the input gain value – 0.0 dB to 111.0 dB. See also NOTES in 4.2.6.
4.4.7
STEP G
Changes the gain step value – 0.1 dB, 0.5 dB, 1.0 dB and 6.0 dB steps.
4.4.8
REJECT
Adjustment of the reject level – all signals below this level will be neither displayed
nor evaluated.
4.4.9
THICKNESS
-
Adjustment of minimum thickness value for automated thickness measurement
using Gate 2 – this value must be lower than any really expected thickness to be
measured.
-
Adjustment of nominal thickness line in the B-scan thickness measuring mode.
4.4.10 VELOCITY
Adjustment of the sound velocity value in the tested material – 1 m/s to 19999 m/s.
For details see 4.1.11. VELOCITY.