Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
38/102
4.5 BSCAN Menu
Activates the B-Scan mode.
-
The diagram is always drawn from initial
position from the right to the left. Total
380 last values are displayed.
-
Measured values (saved to the file and
sent via serial line) have co-ordinates
recalculated to 0.1 mm or 0.01 in.
-
When DGS/DAC is switched ON, BSCAN
shows position of the echo over the reference curve.
BSCAN Menu Items on the Screen – Left Side:
4.5.1.1 GATE >--<
Adjustment of the gate start position (without changing its width) – actual gate
number (activated in GATES menu – see 4.4) is indicated in parentheses next to the
gate start position value – this item is displayed by the corresponding gate colour.
4.5.1.2 GATE --<
Adjustment of the gate end position (changes its width) – actual gate number
(activated in GATES menu – see 4.4) is indicated in parentheses next to the gate
end position value.
4.5.1.3 GATE ^
Adjustment of the gate vertical position – actual gate number (activated in GATES
menu – see 4.4) is indicated in parentheses next to the gate vertical position value.
4.5.1.4 TR MODE
Switches the transmitter/receiver modes – see 4.1.4 for details.
4.5.1.5 PROBE DELAY
Adjustment of probe delay line value (e.g. for dual element or angle probe) within the
range from -10.00
µ
s to 4800.00
µ
s.