Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
14/102
2.1 Connectors and Communication Ports
Note: We strongly recommend to use original cables from DRC for all
connectors.
T – Transmitter (Pulser).
R – Receiver.
ANALOG OUT (for details see 6. Functions of Analog Output):
- TTLout: - synchronization pulses output with positive
logic; can be used for generator
synchronization during calibration, or for
synchronization of two D1000 instruments;
- AOUT2: - voltage 0-BSCAN_MAX as measured
thickness with Gate2: working, voltage
mapping is made by BSCAN MAX value;
- logic info about echo presence within Gate1,
when Gate1 is set to to ON+A/TTL;
- amplitude of echo in Gate1, when Gate1 is set
to ON+AUX and Gate2 is OFF;
- AOUT1: - voltage 0-2.7V as relative info about position
of echo flank in Gate1 or in Gate4 with
EcAUX;
- TxD:
- transmitted data from D1000 (pin 2 on
CANON9F);
- RxD:
- received data (pin 3 on CANON9F);
- GND:
- ground (pin 5 on CANON9F).
MUX
– for future options.
MULTI – primarily for Encoder (IRC) connection,
servers as SYNC input.
USB
POWER
General
Dimensions
224×188×37 mm
Display
Color TFT 99×30 mm (1024×768 pixels)
65535 colors
Weight
0.74 Kg without battery
+ 0.54 kg battery for 8 hours operation
Operating Temperature
-10 °C to 50 °C
Storage Temperature
-40 °C to 70 °C
Power
100-240 VAC, 50 to 60 Hz, 50 VA max. for battery
charging and operation
Battery
Built-in rechargeable Li-Ion battery pack,
3.6 V, 16 Ah
GND
+5V
C
A
B
GND
TTLout
AOUT2
AOUT1
R x D
T x D